LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.
Recall
- Recall Number
- Z-1615-2024
- Event Number
- 94258
- Firm
- XENEX Disinfection Services Inc.
- FEI Number
- 3014531752
- Product Code
- QXJ
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- March 18, 2024
- Posted
- April 23, 2024
- Address
- 1074 Arion Cir, Ste 116, San Antonio, TX, 78216-3085
Description
LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.
Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.
On 3/18/24, correction notices were emailed to customers informing them of the following: 1) Blackout Curtains can continue to be used to block nuisance light coming from an unoccupied room/area where a UV cycle is being executed. It is not necessary to return curtains. 2) If your practice is to use this Blackout Curtain as a protective barrier, discontinue the use in occupied spaces. 3) If you may have further distributed this product, please identify your customers, and notify them at once of this product field correction or provide us with their contact information so the firm can reach out to them. Complete and return the response form. If you have any questions, call the firm at (830) 832-4299 between the hours of 8:00AM and 5:00PM CST Monday through Friday.
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.
377