Maquet Getinge-BEQ-T 25501 3/8" Better Bladder Pack Material: 709000370
Recall
- Recall Number
- Z-1613-2019
- Event Number
- 82703
- Firm
- Maquet Cardiovascular, LLC
- FEI Number
- 2242352
- Product Code
- DWF
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 18, 2019
- Posted
- May 31, 2019
- Terminated
- July 28, 2025
- Address
- 45 Barbour Pond Dr, Wayne, NJ, 07470-2094
Description
Maquet Getinge-BEQ-T 25501 3/8" Better Bladder Pack Material: 709000370
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051.
Nationwide