Galil Medical, 17G Cryoablation Needle Holder FPRPR 4014, Rx only, Sterile EO. Product Usage: Galil Medicals 17G Cryoablation Needle Holder (Needle Holder) is intended for use only with a Galil Medical Needle Testing Device (NTD). The NTD, a component used when performing cryogenic destruction of tissue with a Galil Medical Cryoablation System, is intended for performing a Needle Integrity and Functionality Test in preparation for a cryoablation procedure, and also serves to organize and support the cryoablation needles and thermal sensors in a sterile environment, prior to use.
Recall
- Recall Number
- Z-1613-2012
- Event Number
- 61657
- Firm
- Galil Medical, Inc.
- FEI Number
- 3004462490
- Product Code
- GEH
- Status
- Terminated
- Root Cause
- Material/Component Contamination
- Initiated
- April 12, 2012
- Posted
- May 16, 2012
- Terminated
- August 16, 2012
- Address
- 4364 Round Lake Rd, W Arden Hills, MN, 55112
Description
Galil Medical, 17G Cryoablation Needle Holder FPRPR 4014, Rx only, Sterile EO. Product Usage: Galil Medicals 17G Cryoablation Needle Holder (Needle Holder) is intended for use only with a Galil Medical Needle Testing Device (NTD). The NTD, a component used when performing cryogenic destruction of tissue with a Galil Medical Cryoablation System, is intended for performing a Needle Integrity and Functionality Test in preparation for a cryoablation procedure, and also serves to organize and support the cryoablation needles and thermal sensors in a sterile environment, prior to use.
Galil Medical is recalling 17G Cryoablation Needle Holder, FPRPR 4014. This voluntary recall has been initiated due to the possibility of excessive bioburden on the Needle Holders.
Galil Medical sent an "Urgent Medical Device Recall" letter dated April 12, 2012 to all affected customers. The letter identified the affected product, problem, actions to be taken. The letter instructs customers to return all unused affected products, use a revised needle preparation procedure and forward and communicate this Recall Notice and instructions. Customers were instructed to complete and return the attached Response Form with returned product or fax to 1+877-510-7757. For any questions contact Galil Medical Customer Service.
Worldwide Distribution - USA (nationwide) including the states of CA, NY, NC, WA and the countries of CANADA, DENMARK, FRANCE, NETHERLANDS, TAIWAN and UNITED KINGDOM.
26 kits