FDA Recall Terminated

5.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.

Recall: Z-1612-2010 · Initiated April 21, 2010

Recall

Recall Number
Z-1612-2010
Event Number
55508
Firm
Blackstone Medical, Inc.
FEI Number
3001238051
Product Code
HWX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 21, 2010
Posted
May 19, 2010
Terminated
August 10, 2010
Address
1211 Hamburg Tpke, Suite 300, Wayne, NJ, 07470

Description

5.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.

Reason

Certain lots of the non-cannulated Modular Bone Taps used with the Firebird Spinal Fixation Systems may contain a manufacturing non conformance resulting in the taps creating a single lead screw thread rather than the required dual lead screw thread.

Action

Orthofix issued an "Urgent Product Recall" notification dated April 21, 2010 by USPS certified mail to hospitals and distributors. Consignees were asked to return all affected product to the firm. For additional information, contact the Orthofix Regulatory Affairs Department at 1-973-406-2847.

Distribution

Nationwide and to one location each in Australia and Spain.

Quantity

94