Advantx Legacy Radiographic and Fluoroscopic systems, model 2102424, 2118104, 2157009, 2183000, 2206038, 2206048, 2206056, 2224559, 2258627, 2268730, 2330340, 2333974, 2118104-2, 2118104-3, 2118104-4, 2118104-5, 2118104-6, 2118104-7, 2118104-8, 2139574-2, 2147592-13, 2171204-2, 46-262751G6, 46-262751G7, 46-262751G8, 46-275382G58, and 46-302056G1. GE Medical Systems - Americas: Milwaukee, USA. The device is used to perform fluoroscopic examination of human anatomy.
Recall
- Recall Number
- Z-1611-2008
- Event Number
- 49467
- Firm
- GE Healthcare
- FEI Number
- 2126677
- Product Code
- IXR
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 2, 2008
- Posted
- September 19, 2008
- Terminated
- December 10, 2011
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Advantx Legacy Radiographic and Fluoroscopic systems, model 2102424, 2118104, 2157009, 2183000, 2206038, 2206048, 2206056, 2224559, 2258627, 2268730, 2330340, 2333974, 2118104-2, 2118104-3, 2118104-4, 2118104-5, 2118104-6, 2118104-7, 2118104-8, 2139574-2, 2147592-13, 2171204-2, 46-262751G6, 46-262751G7, 46-262751G8, 46-275382G58, and 46-302056G1. GE Medical Systems - Americas: Milwaukee, USA. The device is used to perform fluoroscopic examination of human anatomy.
GE Healthcare was notified that 2 recently installed Precision 500D systems were not in compliance with portions of the Performance Standard for Diagnostic X-ray Equipment. X-ray production was possible from the fluoroscopic X-ray tube when the Primary Protective Barrier was not in position to intercept the X-ray beam. Users are recommended to follow good clinical standards to ensure the Image I
A Product Safety Notification letter, dated 01/09/2008 was sent to affected customers. This letter described the safety issue, affected product and safety instructions. A GE Field Engineer will schedule a field visit to affected customers to inspect and replace the park latch switch as needed.
Nationwide, Washington D.C. and Puerto Rico and OUS to include: ARGENTINA, AUSTRALIA, BRAZIL, CANADA, CHILE, DOMINICAN REPUBLIC, EGYPT, GERMANY, HONDURAS, INDIA, INDONESIA, IRELAND, JAMAICA, KOREA, KUWAIT, MALAYSIA, MEXICO, NETHERLANDS, PANAMA, RUSSIAN FEDERATION, SAUDIA ARABIA, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM,
2863 total (2744 US / 119 OUS)