cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Recall
- Recall Number
- Z-1610-2018
- Event Number
- 79676
- Firm
- Roche Diagnostics Corporation
- FEI Number
- 1823260
- Product Code
- CEM
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 22, 2018
- Posted
- May 3, 2018
- Terminated
- November 15, 2019
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
ROCHE sent an Urgent Medical Device Correction Letter dated February 22, 2018, via UPS to affected customers. Customers were instructed to complete the following: " 1) Use the Return to Setup and Exit buttons to ensure execution of scheduled AutoQC measurements as described in this Urgent Medical Device Correction (UMDC); 2) Check to see if the next scheduled AutoQC measurement is performed to ensure the AutoQC scheduler is working as expected; 3) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site; 4) Complete the attached fax form (TP-00146) and fax it to 1-844-491-7776 or scan and email to [email protected]; 5) File this UMDC for future reference. Customers with questions should contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272.
USA ( nationwide)
1026 total products