FDA Recall Terminated

Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

Recall: Z-1608-2015 · Initiated May 8, 2015

Recall

Recall Number
Z-1608-2015
Event Number
71233
Firm
Spacelabs Healthcare Inc
FEI Number
3010157426
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
May 8, 2015
Posted
May 13, 2015
Terminated
November 5, 2015
Address
35301 SE Center St, Snoqualmie, WA, 98065-9216

Description

Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

Reason

The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).

Action

The firm sent the Urgent Medical Device Correction- Spacelabs Healthcare Inc.-qube Compact Monitor, Model 91390, Power Input PCBA Failure letter, dated 8 May 2015, via priority service mail to US consignees. This letter described the affected system, problem, actions to be taken by users and corrective actions conducted by Spacelabs. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support

Distribution

distributed in North Carolina, Texas, and Wyoming

Quantity

5 units distributed in the US