Various Alcon Custom Paks Surgical Procedure Pack containing Oasis 27Ga Chang Hydrodissection Cannulas. These Custom Paks are surgical procedure packs used by ophthalmic surgeons in a variety of ophthalmic surgeries. The Custom Pak is a collection of various ophthalmic surgical instruments and necessary materials. Below are the various Alcon Customer Paks and Pak REF Numbers: 0515-66 ELMENDORF AFB CATARACT 3RD MED CTR 0752-74 CATARACT V A MED CTR 1119-82 DR DAVID B LEACH PALOUSE SURGERY CENTER 11401-13 CATARACT SUTTER SOLANO MED CTR 11910-32 DAVISON-SAGGAU CATARACT WOLFE EYE SURGER 11912-26 30 DEG CATARACT BASE PACK WOLFE EYE SUR 11913-23 JOHNSON CATARACT WOLFE EYE SURGERY CENTE 12185-14 CATARACT LASERVUE EYE CENTER 12213-12 CATARACT TULSA SPINE AND SPECIALTY 12213-13 CATARACT TULSA SPINE AND SPECIALTY 12213-15 CATARACT TULSA SPINE AND SPECIALTY 12213-16 CATARACT TULSA SPINE AND SPECIALTY 12643-06 CATARACT PACK OCHSNER BAPTIST MED CTR 12931-17 JOHNSON CATARACT GREATER COMMUNITY HOSP 12942-18 CONRAD PHYSICIANS MEDICAL CENTER 12942-19 CONRAD PHYSICIANS MEDICAL CENTER 13567-10 OPHTHALMIC PHYSICIANS SURG CTR 13948-15 DR RHEE PACK MARSHALLTOWN MED/SURG CNT 13970-11 CATARACT GWINNETT CTR OUTPT SURG 13970-12 CATARACT GWINNETT CTR OUTPT SURG 14389-08 INFINITI STAMFORD SURGICAL CENTER 14408-12 E-Z PACK QUEEN OF THE VALLEY 14408-13 E-Z PACK QUEEN OF THE VALLEY 14533-18 OPHTHALMIC UOFL HOSPITAL 15209-05 CATARACT NORTH ATLANTA EYE SURGERY CTR 15354-16 DR RAUEN WOLFE EYE SURGERY CENTER LLC 15435-13 EPS 2.2 PACK IOWA CITY AMBULATORY SURGER 15502-10 DR NEWCOMB LEXINGTON CLINIC 15639-12 DR RAUEN LUCAS COUNTY HEALTH CTR 15678-13 DR RAUEN CLARKE COUNTY HOSPITAL 15965-12 DR LEVAR CHRISTUS SANTA ROSA ASC 16001-12 DR YONKER/BESCAK CESC 16001-13 DR YONKER/BESCAK CESC 16013-12 DR MCDONALD NORTH FL SURG PAVILION 16086-07 DR DAVID B LEACH PULLMAN MEMORIAL HOSPIT 16177-08 OPHTHALMIC DENVER EYE SURG CTR 16204-13 CATARACT V A HOSPITAL 16450-10 DR GROSS/GUSSLER KINGS DAUGHTERS HOSPITA 16452-06 RAECKER PACK GREATER COMMUNITY HOSP 16633-08 CATARACT DUOVISC RETINA SPECIALISTS NJ 16644-08 DR BALES HOSPITAL PACK NATIVE INVESTMENT 16644-09 DR BALES HOSPITAL PACK NATIVE INVESTMENT 16798-08 CATARACT HAUSER ROSS EYE INSTITUTE 16820-06 PHACO LONE PEAK SURGERY CENTER 17029-09 DR ICON CATARACT DAY SURGERY OF GRAND JU 17201-18 SILVERA VALLEY SURGERY CENTER 17340-08 EPS 2.2 PACK WASHINGTON COUNTY HOSP 17354-07 EPS 2.2 PACK HENRY COUNTY HEALTH CTR 17361-08 MICRO PAK CARLE FOUNDATION HOSP 17382-05 CATARACT BILLINGS VA MEDICAL CENTE 17382-07 CATARACT BILLINGS VA MEDICAL CENTE 17507-02 DR CAMPAGNA NORTH CENTRAL METH ASC 17520-05 DR GROSS CHARLESTON SURGERY CTR 17526-09 CPK0926-CROSBY-DONG-GADLIN-JAC VANTAGE O 17529-09 CPK0927 AESC VANTAGE OUTSOURCING 17541-07 HAIDER DUPONT SURG CTR 17541-08 HAIDER DUPONT SURG CTR 17617-05 DR CHUNG CATARACT PACK ST HELENA HOSPITA 17651-06 DMEK-DSEK-CAT CENTER FOR ADVANCED SURGER 17653-06 DR HUNT CHARLESTON AREA MED CTR 17671-06 CATARACT EDMONDS CENTER FOR OUTPAT 17671-07 CATARACT EDMONDS CENTER FOR OUTPAT 17740-06 DR ESPANDAR ARMSTRONG CO MEM HOSP 17740-07 DR ESPANDAR ARMSTRONG CO MEM HOSP 17740-08 DR ESPANDAR ARMSTRONG CO MEM HOSP 17773-09 DR ESPANDAR JAMESON MEMORIAL HOSPITAL 17930-08 CATARACT PACK NORTHERN PLAINS SURG CTR 17966-06 TURNER CATARACT WOLFE EYE SURGERY CENTER 17987-05 OPHTHALMIC SEATTLE SURGERY CENTER 18069-04 DR WENTWORTH CATARACT GREENVIEW SURGERY 18123-05 CATARACT BOWERS INNOVATIVE EYE SURGERY C 18208-09 CATARACT SUTTER FAIRFIELD SURG CTR 18382-04 DR GEORGE ZAMBELLI BEAVER VALLEY ASC 18382-06 DR GEORGE ZAMBELLI BEAVER VALLEY ASC 18384-04 DR BURKS SPRINGHILL SURGERY CENTER 18401-04 MCGRAW SIGHTPATH MEDICAL 18406-03 DR KUCHINKA HAYWARD AREA HOSP 18427-07 CATARACT V A HOSPITAL 18427-08 CATARACT V A HOSPITAL 18431-06 VISION EYE SC VISION EYE SURGERY CENTER 18445-03 K LEE/LIN CALIFORNIA PACIFIC MED CT 18550-04 DR HU TORRES CAT & LSR CNTRL 18555-04 DR WU - CENTURION CAT & LSR CNTRL 18636-10 CATARACT BONNER GENERAL HOSPITAL
Recall
- Recall Number
- Z-1606-2024
- Event Number
- 94253
- Firm
- Alcon Research, LLC
- FEI Number
- 1000117501
- Product Code
- LRO
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- March 15, 2024
- Posted
- April 22, 2024
- Address
- 9965 Buffalo, Speedway Houston, TX, 77054-1309
Description
Various Alcon Custom Paks Surgical Procedure Pack containing Oasis 27Ga Chang Hydrodissection Cannulas. These Custom Paks are surgical procedure packs used by ophthalmic surgeons in a variety of ophthalmic surgeries. The Custom Pak is a collection of various ophthalmic surgical instruments and necessary materials. Below are the various Alcon Customer Paks and Pak REF Numbers: 0515-66 ELMENDORF AFB CATARACT 3RD MED CTR 0752-74 CATARACT V A MED CTR 1119-82 DR DAVID B LEACH PALOUSE SURGERY CENTER 11401-13 CATARACT SUTTER SOLANO MED CTR 11910-32 DAVISON-SAGGAU CATARACT WOLFE EYE SURGER 11912-26 30 DEG CATARACT BASE PACK WOLFE EYE SUR 11913-23 JOHNSON CATARACT WOLFE EYE SURGERY CENTE 12185-14 CATARACT LASERVUE EYE CENTER 12213-12 CATARACT TULSA SPINE AND SPECIALTY 12213-13 CATARACT TULSA SPINE AND SPECIALTY 12213-15 CATARACT TULSA SPINE AND SPECIALTY 12213-16 CATARACT TULSA SPINE AND SPECIALTY 12643-06 CATARACT PACK OCHSNER BAPTIST MED CTR 12931-17 JOHNSON CATARACT GREATER COMMUNITY HOSP 12942-18 CONRAD PHYSICIANS MEDICAL CENTER 12942-19 CONRAD PHYSICIANS MEDICAL CENTER 13567-10 OPHTHALMIC PHYSICIANS SURG CTR 13948-15 DR RHEE PACK MARSHALLTOWN MED/SURG CNT 13970-11 CATARACT GWINNETT CTR OUTPT SURG 13970-12 CATARACT GWINNETT CTR OUTPT SURG 14389-08 INFINITI STAMFORD SURGICAL CENTER 14408-12 E-Z PACK QUEEN OF THE VALLEY 14408-13 E-Z PACK QUEEN OF THE VALLEY 14533-18 OPHTHALMIC UOFL HOSPITAL 15209-05 CATARACT NORTH ATLANTA EYE SURGERY CTR 15354-16 DR RAUEN WOLFE EYE SURGERY CENTER LLC 15435-13 EPS 2.2 PACK IOWA CITY AMBULATORY SURGER 15502-10 DR NEWCOMB LEXINGTON CLINIC 15639-12 DR RAUEN LUCAS COUNTY HEALTH CTR 15678-13 DR RAUEN CLARKE COUNTY HOSPITAL 15965-12 DR LEVAR CHRISTUS SANTA ROSA ASC 16001-12 DR YONKER/BESCAK CESC 16001-13 DR YONKER/BESCAK CESC 16013-12 DR MCDONALD NORTH FL SURG PAVILION 16086-07 DR DAVID B LEACH PULLMAN MEMORIAL HOSPIT 16177-08 OPHTHALMIC DENVER EYE SURG CTR 16204-13 CATARACT V A HOSPITAL 16450-10 DR GROSS/GUSSLER KINGS DAUGHTERS HOSPITA 16452-06 RAECKER PACK GREATER COMMUNITY HOSP 16633-08 CATARACT DUOVISC RETINA SPECIALISTS NJ 16644-08 DR BALES HOSPITAL PACK NATIVE INVESTMENT 16644-09 DR BALES HOSPITAL PACK NATIVE INVESTMENT 16798-08 CATARACT HAUSER ROSS EYE INSTITUTE 16820-06 PHACO LONE PEAK SURGERY CENTER 17029-09 DR ICON CATARACT DAY SURGERY OF GRAND JU 17201-18 SILVERA VALLEY SURGERY CENTER 17340-08 EPS 2.2 PACK WASHINGTON COUNTY HOSP 17354-07 EPS 2.2 PACK HENRY COUNTY HEALTH CTR 17361-08 MICRO PAK CARLE FOUNDATION HOSP 17382-05 CATARACT BILLINGS VA MEDICAL CENTE 17382-07 CATARACT BILLINGS VA MEDICAL CENTE 17507-02 DR CAMPAGNA NORTH CENTRAL METH ASC 17520-05 DR GROSS CHARLESTON SURGERY CTR 17526-09 CPK0926-CROSBY-DONG-GADLIN-JAC VANTAGE O 17529-09 CPK0927 AESC VANTAGE OUTSOURCING 17541-07 HAIDER DUPONT SURG CTR 17541-08 HAIDER DUPONT SURG CTR 17617-05 DR CHUNG CATARACT PACK ST HELENA HOSPITA 17651-06 DMEK-DSEK-CAT CENTER FOR ADVANCED SURGER 17653-06 DR HUNT CHARLESTON AREA MED CTR 17671-06 CATARACT EDMONDS CENTER FOR OUTPAT 17671-07 CATARACT EDMONDS CENTER FOR OUTPAT 17740-06 DR ESPANDAR ARMSTRONG CO MEM HOSP 17740-07 DR ESPANDAR ARMSTRONG CO MEM HOSP 17740-08 DR ESPANDAR ARMSTRONG CO MEM HOSP 17773-09 DR ESPANDAR JAMESON MEMORIAL HOSPITAL 17930-08 CATARACT PACK NORTHERN PLAINS SURG CTR 17966-06 TURNER CATARACT WOLFE EYE SURGERY CENTER 17987-05 OPHTHALMIC SEATTLE SURGERY CENTER 18069-04 DR WENTWORTH CATARACT GREENVIEW SURGERY 18123-05 CATARACT BOWERS INNOVATIVE EYE SURGERY C 18208-09 CATARACT SUTTER FAIRFIELD SURG CTR 18382-04 DR GEORGE ZAMBELLI BEAVER VALLEY ASC 18382-06 DR GEORGE ZAMBELLI BEAVER VALLEY ASC 18384-04 DR BURKS SPRINGHILL SURGERY CENTER 18401-04 MCGRAW SIGHTPATH MEDICAL 18406-03 DR KUCHINKA HAYWARD AREA HOSP 18427-07 CATARACT V A HOSPITAL 18427-08 CATARACT V A HOSPITAL 18431-06 VISION EYE SC VISION EYE SURGERY CENTER 18445-03 K LEE/LIN CALIFORNIA PACIFIC MED CT 18550-04 DR HU TORRES CAT & LSR CNTRL 18555-04 DR WU - CENTURION CAT & LSR CNTRL 18636-10 CATARACT BONNER GENERAL HOSPITAL
Due to complaints of particulates, residue, occlusions, and bent tips for sterile cannulas.
On 03/15/2024, the firm sent an "URGENT: VOLUNTARY MEDICAL DEVICE FIELD CORRECTION" Letter via FedEx overnight to inform customer that, specific lots of Alcon Custom Pak containing Oasis Medical 27G Chang Hydrodissection Cannulas. Alcon has received customer complaints for foreign material, occlusions and bent tips for these cannulas. We are asking that you remove and dispose of the included 27G Chang Hydrodissection Cannula upon opening your Custom Pak for its specific surgical case. Customer are to take the following actions: 1. See Table 1 for list of affected Alcon Custom Pak lots that were distributed to their facility. They may have received Custom Pak lots assembled prior to this Field Correction notification, and as a result, they may be receiving affected inventory in future shipments. Any affected Alcon Custom Pak in the future, will be identified by a sticker and reads, "URGENT FIELD CORRECTION Discard enclosed 27G Chang Hydrodissection Cannula" 2. Review their inventory of Alcon Custom Pak lots against the lots listed in Table 1. For affected lots in their current on-hand inventory, affix the provided stickers directly to the outside of their affected Custom Pak in a location most easily seen. NOTE: Alcon Custom Pak units are sterile and should not be opened prior to surgery. 3. Upon opening of the identified Alcon Custom Pak for surgical use, remove and dispose of the enclosed 27G Chang Hydrodissection Cannula. 4. Use a separate sterile standalone hydrodissection cannula, either from their own inventory or those supplied by Alcon. For impacted products that will ship in the future, Alcon is planning to ship sterile stand-alone substitutions. Call Alcon Customer Service (1-800-862-5266) if help it needed in obtaining replacement sterile standalone hydrodissection cannulas. 5. Keep Table 1 for their records should they receive any affected Alcon Custom Pak lots in the near future. Questions/concerns, contact Alcon Customer Service at 1-800-862-5266.
Worldwide distribution - U.S. Nationwide including in the states of AK, AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Australia, Canada, and New Zealand.
57,352 cannulas