FDA Recall Terminated

GE Healthcare, CARESCAPE Monitor B850, B650 and B450 (Bx50).

Recall: Z-1606-2015 · Initiated May 4, 2015

Recall

Recall Number
Z-1606-2015
Event Number
71211
Firm
GE Healthcare
FEI Number
2126677
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
May 4, 2015
Posted
May 12, 2015
Terminated
September 15, 2015
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare, CARESCAPE Monitor B850, B650 and B450 (Bx50).

Reason

The heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message and there may not be an alarm for asystole. Undetected asystole could result in irreversible changes in the patients condition and delayed or missed life sustaining patient treatment

Action

GE Healthcare sent an Urgent Medical Device Correction letter dated May 1, 2015, to all affected consignees. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution - US including CO, MI, WI and Internationally to NETHERLANDS, FINLAND, and BELGIUM.

Quantity

110 (83 units US, 27 units OUS)