Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714
Recall
- Recall Number
- Z-1606-2008
- Event Number
- 47518
- Firm
- Siemens Healthcare Diagnostics Inc.
- Product Code
- MTF
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 24, 2008
- Posted
- June 28, 2008
- Terminated
- October 15, 2008
- Address
- 500 GBC Drive P. O. Box 6101, Mailstop 514, Newark, DE, 19714-6101
Description
Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714
False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.
The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.
The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE
1631 cartons