Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714

Recall: Z-1606-2008 · Initiated March 24, 2008

Recall

Recall Number
Z-1606-2008
Event Number
47518
Firm
Siemens Healthcare Diagnostics Inc.
Product Code
MTF
Status
Terminated
Root Cause
Device Design
Initiated
March 24, 2008
Posted
June 28, 2008
Terminated
October 15, 2008
Address
500 GBC Drive P. O. Box 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714

Reason

False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.

Action

The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.

Distribution

The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE

Quantity

1631 cartons