Nuvectra, Algovita Spinal Cord Stimulation System Clinical Programmer, Model 4500. For use with Algovita Spinal Cord Stimulation Systems. The version of the software in the Clinician Programmer is v1.1.5. The Algovita Spinal Cord Stimulation system consists of a stimulator (EPG or IPG) that is physically and electrically connected to one or more leads inserted into the patient"s spinal epidural space. The lead delivers electrical stimulation originating at the stimulator with the purpose of blocking pain signals going to the patient"s brain. The stimulators are rechargeable, and the frequency of recharging is dependent on individual patient"s use of the system. The Clinician Programmer (CP) is a hand-held, touch screen device used to create and adjust stimulation parameters that are developed to optimize the therapy for each patient. The CP communicates with the stimulator via the Medical Implant Communication Service (MICS). The CP wirelessly transfers the stimulation parameters to the stimulator. In most clinical settings, the CP can be used by the Health Care Professional to program multiple patient EPGs and IPGs

Recall: Z-1600-2016 · Initiated April 6, 2016

Recall

Recall Number
Z-1600-2016
Event Number
73884
Firm
Nuvectra
FEI Number
3010309840
Product Code
LGW
Status
Terminated
Root Cause
Software design
Initiated
April 6, 2016
Posted
May 3, 2016
Terminated
September 29, 2017
Address
10675 Naples St NE, Blaine, MN, 55449-5802

Description

Nuvectra, Algovita Spinal Cord Stimulation System Clinical Programmer, Model 4500. For use with Algovita Spinal Cord Stimulation Systems. The version of the software in the Clinician Programmer is v1.1.5. The Algovita Spinal Cord Stimulation system consists of a stimulator (EPG or IPG) that is physically and electrically connected to one or more leads inserted into the patient"s spinal epidural space. The lead delivers electrical stimulation originating at the stimulator with the purpose of blocking pain signals going to the patient"s brain. The stimulators are rechargeable, and the frequency of recharging is dependent on individual patient"s use of the system. The Clinician Programmer (CP) is a hand-held, touch screen device used to create and adjust stimulation parameters that are developed to optimize the therapy for each patient. The CP communicates with the stimulator via the Medical Implant Communication Service (MICS). The CP wirelessly transfers the stimulation parameters to the stimulator. In most clinical settings, the CP can be used by the Health Care Professional to program multiple patient EPGs and IPGs

Reason

If used, the optional Swap feature copies programming parameters from the external pulse generator (EPG) as well as unintended calibration data to the implantable pulse generator (lPG). This may result in the user being unable to recharge the lPG. This safety notice only applies to CPs Model 4500.

Action

Nuvectra sent an "Urgent Field Safety Notice" letter dated April 6, 2016 to customers. The letter identified the affected product, problem, recommendations and Mitigations, and actions to be taken.