FDA Recall Terminated

Atellica CH Ethyl Alcohol, Test Code ETOH, Siemens Material Number (SMN) 11097501, UDI 00630414596112. The product is an ethyl alcohol assay, IVD use.

Recall: Z-1599-2020 · Initiated February 14, 2020

Recall

Recall Number
Z-1599-2020
Event Number
85020
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
DIC
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
February 14, 2020
Posted
February 27, 2020
Terminated
February 22, 2024
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica CH Ethyl Alcohol, Test Code ETOH, Siemens Material Number (SMN) 11097501, UDI 00630414596112. The product is an ethyl alcohol assay, IVD use.

Reason

The assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.

Action

On February 11, 2020, the firm issued an Urgent Medical Device Correction to all US customers. Customers who experience the issue were instructed to perform a pack calibration every 2 days or more frequently based on the laboratory's Quality Control performance. The firm is working to address the issue in a future software version. Customers are requested to complete and return the effectiveness check questionnaire. If you have any questions, contact your local Siemens Healthineers technical support representative.

Distribution

Domestic distribution nationwide. Foreign distribution to Australia, Austria, Bahamas, Belgium, Canada, Curacao, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kenya, Latvia, Mexico, Netherlands, Norway, Poland, Portugal, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., Uruguay, United Kingdom, and Vietnam.

Quantity

1124