Atellica CH Ethyl Alcohol, Test Code ETOH, Siemens Material Number (SMN) 11097501, UDI 00630414596112. The product is an ethyl alcohol assay, IVD use.
Recall
- Recall Number
- Z-1599-2020
- Event Number
- 85020
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2432235
- Product Code
- DIC
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- February 14, 2020
- Posted
- February 27, 2020
- Terminated
- February 22, 2024
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
Atellica CH Ethyl Alcohol, Test Code ETOH, Siemens Material Number (SMN) 11097501, UDI 00630414596112. The product is an ethyl alcohol assay, IVD use.
The assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.
On February 11, 2020, the firm issued an Urgent Medical Device Correction to all US customers. Customers who experience the issue were instructed to perform a pack calibration every 2 days or more frequently based on the laboratory's Quality Control performance. The firm is working to address the issue in a future software version. Customers are requested to complete and return the effectiveness check questionnaire. If you have any questions, contact your local Siemens Healthineers technical support representative.
Domestic distribution nationwide. Foreign distribution to Australia, Austria, Bahamas, Belgium, Canada, Curacao, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kenya, Latvia, Mexico, Netherlands, Norway, Poland, Portugal, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., Uruguay, United Kingdom, and Vietnam.