FDA Recall Open, Classified

Strep test: a. McKesson Medical-Surgical McKesson Rapid Test Kit McKesson Consult Infectious Disease Immunoassay Strep A Test Throat / Tonsil Saliva Sample 25 Tests, Model Number: 5003. b. QUIDEL Sofia TEST KIT, STREP A+ SOFIA (25/KT 12KT/CS) Model Number: 20274. c. Alere SAN DIEGO INC McKesson TEST, KIT STREP A DIPSTICK WAIVED (50/KT) Exempt Catalog #4999. d. In-Line Control Swab Quickvue In-Line Strep A Test Positive Level / Negative Level 12 Swabs (IVD) Catalog # 00345. e. (added 11/30/22) Abbott Rapid Test Kit Acceava Infectious Disease Immunoassay Strep A Test Throat Swab Sample 50 Tests, Model Number: 92001

Recall: Z-1598-2022 · Initiated May 25, 2022

Recall

Recall Number
Z-1598-2022
Event Number
90336
Firm
Mckesson Medical-Surgical Inc. Corporate Office 9954 Maryland Drive Deep Run Iii Ste. 4000 Richmond VA 23233
FEI Number
3004386693
Product Code
GTY
Status
Open, Classified
Root Cause
Environmental control
Initiated
May 25, 2022
Posted
August 26, 2022

Description

Strep test: a. McKesson Medical-Surgical McKesson Rapid Test Kit McKesson Consult Infectious Disease Immunoassay Strep A Test Throat / Tonsil Saliva Sample 25 Tests, Model Number: 5003. b. QUIDEL Sofia TEST KIT, STREP A+ SOFIA (25/KT 12KT/CS) Model Number: 20274. c. Alere SAN DIEGO INC McKesson TEST, KIT STREP A DIPSTICK WAIVED (50/KT) Exempt Catalog #4999. d. In-Line Control Swab Quickvue In-Line Strep A Test Positive Level / Negative Level 12 Swabs (IVD) Catalog # 00345. e. (added 11/30/22) Abbott Rapid Test Kit Acceava Infectious Disease Immunoassay Strep A Test Throat Swab Sample 50 Tests, Model Number: 92001

Reason

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Action

Mckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 [email protected]. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent possible patient harm. Please complete and return the enclosed response form as soon as possible even if you do not have any product on hand. Customers should contact Sedgwick directly by phone at 1-888-943-2401 to return the recalled products or if they have additional questions regarding this notification. Representatives are available Monday Friday, 8:00 am 5:00 pm PT

Distribution

US Nationwide

Quantity

a. 214 kits b. 1cs c. 53 kits d. 4 bx e. 1 KT