MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Recall
- Recall Number
- Z-1598-2016
- Event Number
- 73947
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Vendor change control
- Initiated
- April 13, 2016
- Posted
- May 3, 2016
- Terminated
- April 17, 2017
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Siemens mailed to customers on April 13, 2016, a Customer Safety Advisory Notice, MR013/16/S, to inform them of the actions to take to prevent a possible hazard to persons and equipment. Customers were informed that they will be contacted by one of Siemens' service organization who will schedule an appointment to perform a correction, which is to exchange the top cover.
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
8 systems