FDA Recall Terminated

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Recall: Z-1598-2016 · Initiated April 13, 2016

Recall

Recall Number
Z-1598-2016
Event Number
73947
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LNH
Status
Terminated
Root Cause
Vendor change control
Initiated
April 13, 2016
Posted
May 3, 2016
Terminated
April 17, 2017
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Reason

The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.

Action

Siemens mailed to customers on April 13, 2016, a Customer Safety Advisory Notice, MR013/16/S, to inform them of the actions to take to prevent a possible hazard to persons and equipment. Customers were informed that they will be contacted by one of Siemens' service organization who will schedule an appointment to perform a correction, which is to exchange the top cover.

Distribution

US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.

Quantity

8 systems