FDA Recall Terminated

TOSHIBA Kalare Diagnostic X-Ray System K110785

Recall: Z-1598-2013 · Initiated April 3, 2013

Recall

Recall Number
Z-1598-2013
Event Number
65049
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
JAA
Status
Terminated
Root Cause
Component design/selection
Initiated
April 3, 2013
Posted
July 10, 2013
Terminated
February 5, 2016
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92780-7047

Description

TOSHIBA Kalare Diagnostic X-Ray System K110785

Reason

Firm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few seconds or more. This occurs when the Automatic Exposure Function (AEC) function is used. As a result, a radiographic tube current exceeding the set value flows and a brighter image than expected is acquired.

Action

Firm initiated recall by sending recall notification letters along with a customer reply form on April 3, 2013 via USPS return mail receipt. A follow-up letter (with customer reply form) was sent on April 11, 2013 to inform customers that they should continue to use the AEC function as usual. The updated letter along with the customer reply form dated April 11, 2013 was also sent via USPS return mail receipt. The letter titled, "URGENT: MEDICAL DEVICE CORRECTION", notified customers of the reason for recall, product description under recall, corrective action, instructions to customer under 1. A new circuit board will be installed in each involved unit by a Toshiba service representative, without any charge to the user. 2. In the meantime while waiting for service, the user is to continue using the device with the AEC function. If the AEC function is not available, the user should push the radiography READY and EXPOSURE switches at the same time. "Note to Customer", and contact information "(800) 421-1968 or your local Toshiba Representative at (800) 521-1968".

Distribution

US Nationwide Distribution

Quantity

153 units