FDA Recall Terminated

Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a variety of screws, hooks, rods, connectors, and staples that are used for attachment to the non-cervical spine. Rods span the distance between screws and hooks and achieve fixation by the mechanical joining of the rods with the screws or hooks. Lordosed rods are pre-curved to better fit the patient anatomy.

Recall: Z-1597-2015 · Initiated April 9, 2015

Recall

Recall Number
Z-1597-2015
Event Number
70950
Firm
Spinal Elements, Inc
FEI Number
3004893332
Product Code
NKB
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 9, 2015
Posted
May 11, 2015
Terminated
June 25, 2015
Address
3115 Melrose Dr, Suite 200, Carlsbad, CA, 92010-6690

Description

Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a variety of screws, hooks, rods, connectors, and staples that are used for attachment to the non-cervical spine. Rods span the distance between screws and hooks and achieve fixation by the mechanical joining of the rods with the screws or hooks. Lordosed rods are pre-curved to better fit the patient anatomy.

Reason

Some rods may not meet the appropriate traceability requirements including the lot number. Some rods were found to not include the product identification number (product number and lot number) marked directly on the rod. The lack of marking interrupts the ability to maintain traceability of the device.

Action

The firm sent customer notification letters on 04/09/15. In this letter the firm requested that all affected lots be returned, and replacement inventory will be made available. The firm requested that customers take the following actions: 1. Immediately examine inventory for part number 60100-045. 2. Immediately discontinue the use and/or distribution of the affected lots 3. Send all affected product back to: Spinal Elements, Inc. ATTN: Julie Lamothe 3115 Melrose Drive Carlsbad, CA 92010 4. If further distributed the product to please identify the customers and notify them as once. 5. Complete and return the enclosed "Customer Response Form." Any questions the firm states to contact the Customer Service department at 1.877.SPINAL5 Ext. 213.

Distribution

Nationwide Distribution including the states of GA, CA, TX, CO, IN, NC, FL, UT, TN, MO, MI, PA, OH, and OK.

Quantity

588