FDA Recall Terminated

ASC2000 Ambulatory Surgical Table. Model #MTS110003. The device is used to provide complete and flexible positioning for surgical procedures.

Recall: Z-1597-2009 · Initiated April 17, 2009

Recall

Recall Number
Z-1597-2009
Event Number
51953
Firm
Steris Corp
FEI Number
1527821
Product Code
GDC
Status
Terminated
Root Cause
Device Design
Initiated
April 17, 2009
Posted
July 7, 2009
Terminated
March 1, 2012
Address
5960 Heisley Rd, Mentor, OH, 44060

Description

ASC2000 Ambulatory Surgical Table. Model #MTS110003. The device is used to provide complete and flexible positioning for surgical procedures.

Reason

Upon investigation, the firm determined that repetitive rubbing of the hydraulic hoses against the interior of the stainless steel shroud may result in the rupture of the hydraulic hoses causing loss of control of the table surface.

Action

A Field Correction Notice dated April 17, 2009 was issued to customers. The letter described the potential problem of the unit. The letter also notified customers of the corrective actions to be taken by the firm. A Steris technician will visit affected accounts to inspect the hydraulic hoses and replace any that show abrasion and will make other preventive corrections. For further information or if you have questions regarding the servicing of your unit, please contact the STERIS Field Service Dispatch at 1-800-288-2805.

Distribution

Worldwide Distribution -- US (states of AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MD, MI, MN, MO, NC, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TN, TX, VA and WV), AUSTRALIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, ECUADOR, GERMANY, HONG KONG, INDIA, INDONESIA, ITALY, KOREA, MEXICO, PHILIPPINES, PORTUGAL, SINGAPORE, SPAIN, SRI LANKA and THAILAND.

Quantity

412 units