FDA Recall Terminated

Alto VR (model 615) and Alto 2 VR(model 625) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Single chamber arrhythmia detection. Antibradycardia pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring ventricular activity to identify persistent ventricular arrhythmias and to deliver appropriate therapies.

Recall: Z-1597-05 · Initiated July 19, 2005

Recall

Recall Number
Z-1597-05
Event Number
32687
Firm
ela Medical Llc
FEI Number
2182863
Product Code
MRM
Status
Terminated
Root Cause
Other
Initiated
July 19, 2005
Posted
September 24, 2005
Terminated
January 28, 2007
Address
2950 Xenium Ln N, Ste 120, Plymouth, MN, 55441-2623

Description

Alto VR (model 615) and Alto 2 VR(model 625) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Single chamber arrhythmia detection. Antibradycardia pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring ventricular activity to identify persistent ventricular arrhythmias and to deliver appropriate therapies.

Reason

Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.

Action

A Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician.

Distribution

Nationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI.

Quantity

51 total devices (1-model 615 & 50-model 625)