FDA Recall
Terminated
Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
Recall: Z-1596-2011
·
Initiated October 27, 2010
Recall
- Recall Number
- Z-1596-2011
- Event Number
- 57892
- Firm
- Hospira Inc
- FEI Number
- 2921482
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- October 27, 2010
- Posted
- March 8, 2011
- Terminated
- November 7, 2013
- Address
- 755 Jarvis Dr, Morgan Hill, CA, 95037-2810
Description
Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
Reason
Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.
Action
Hospira Recall notification letters were sent by Federal Express on November 1, 2010.
Distribution
Nationwide distribution and to Australia, Canada, and Malaysia.
Quantity
32,376 power cords