FDA Recall Terminated

StoneBreaker C02 Cartridge, Catalog number: SBC-10; GPN: G52603; UDI: (01)10827002526034

Recall: Z-1595-2018 · Initiated January 31, 2018

Recall

Recall Number
Z-1595-2018
Event Number
79731
Firm
Cook Inc.
FEI Number
1820334
Product Code
FFK
Status
Terminated
Root Cause
Error in labeling
Initiated
January 31, 2018
Terminated
April 17, 2020
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

StoneBreaker C02 Cartridge, Catalog number: SBC-10; GPN: G52603; UDI: (01)10827002526034

Reason

The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.

Action

On March 8, 2018 Cook Medical sent customers notification of an Urgent Medical Device Recall . Customers were instructed to complete the following actions: 1. Examine inventory immediately to determine if you have affected product, and quarantine affected product(s). Immediately cease all distribution and use of these products. 2.Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form. Used devices being returned should be double-bagged, placed inside an outer puncture free package with sufficient cushioning material to prevent movement between the secondary container and the outer package. An itemized list of the components must be placed in the container and the outer package marked Used Medical Device. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or email ([email protected]). 4. Immediately report adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email to: [email protected]. Customers with medical questions or concerns can contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235

Distribution

US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA

Quantity

38,542