FDA Recall Terminated

TCM CombiM Modules 903-111

Recall: Z-1595-2016 · Initiated November 18, 2015

Recall

Recall Number
Z-1595-2016
Event Number
73601
Firm
Radiometer America Inc
FEI Number
3010420736
Product Code
LKD
Status
Terminated
Root Cause
Software design
Initiated
November 18, 2015
Terminated
March 27, 2018
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

TCM CombiM Modules 903-111

Reason

Radiometer America, Inc. is recalling TCM CombiM on rare occasions they may have an error that can cause the alarm to sound higher than the pre-set maximum volume.

Action

RadioMeter sent an Product Notification letter out on November 18, 2015, January 10, 2016 and March 29, 2016. The notification sent on November 18, 2015 and January 10, 2016, instructed customers a temporary solution of switching off the analyzer and then on again in order to re-establish the pre-set alarm volume. If alarm sound is higher than the pre-set volume customers are instructed to contact their local Radiometer service representative. If any questions contact Radiometer America Technical Support at 1-800-736--600 opt. 4. The notification sent on March 29, 2016, provided the same instructions and temporary solution. The notification also informed customers that the firm will be in contact with them to discuss the upgrade process.

Distribution

Worldwide Distribution - USA (nationwide) Distribution to the states of : MO, NE, AL, PA, IL, MN, TX, ID, CO, UT, NV, CA, NY, RI, OR, NJ, AZ, GA, IN, VA, LA, NM, MD and AR., and to the countries of : Canada, Austria, Czech Republic, Denmark, Finland, France, Germany, Hungary, Ireland, United Kingdom, Italy, Netherlands, Norway, Spain, Sweden, United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Oman, Jordan, Croatia, Malta, Iceland, Brazil, Thailand, Philippines, Peru, Honduras, Mexico, Kenya, Israel, Romania, Russia, Poland, Turkey, Australia, New Zealand, China, Hong Kong, South Africa, India, Korea, Singapore and Taiwan.

Quantity

1488