FDA Recall
Terminated
Basic Core Pack # OR1805, Sterile, Tri State Hospital Supply Corporation Manufacturers of Centurion Healthcare Products.
Recall: Z-1595-2010
·
Initiated April 9, 2010
Recall
- Recall Number
- Z-1595-2010
- Event Number
- 55305
- Firm
- Centurion Medical Products
- FEI Number
- 1824619
- Product Code
- GEI
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 9, 2010
- Posted
- May 12, 2010
- Terminated
- October 25, 2010
- Address
- 301 Catrell Dr, Howell, MI, 48843-1703
Description
Basic Core Pack # OR1805, Sterile, Tri State Hospital Supply Corporation Manufacturers of Centurion Healthcare Products.
Reason
The firm is conducting a sub recall of kits containing ConMed Goldline Rocker Switch Pencils. The pencils activation switch may remain in the "on" position after the switch has been depressed with excessive force, thereby continuously activating the pencil.
Action
Centurion sent out an "Urgent Product Recall Notice" to one customer on 4/9/2010. The customer was instructed to discontinue use of the product. The customer was also instructed to return any instock product .
Distribution
TX
Quantity
74 kits