FDA Recall Terminated

Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.

Recall: Z-1592-05 · Initiated August 17, 2005

Recall

Recall Number
Z-1592-05
Event Number
33032
Firm
Baxter Healthcare Corp. Rt.
FEI Number
1416980
Product Code
FRN
Status
Terminated
Root Cause
Other
Initiated
August 17, 2005
Posted
September 28, 2005
Terminated
June 19, 2007
Address
120 & Wilson Rd, Round Lake, IL, 60073

Description

Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.

Reason

Potential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher

Action

Baxter sent the 8/17/05 Urgent Product Recall letter to all Baxter 6060E Infusion Pump customers to inform them of a software anomaly present in software versions 3.00 and higher, which allows the pump to overinfuse in the Auto-Ramp mode. The end users were also informed that these epidural pumps, identified by the yellow housing, can no longer be supported due to a lack of availability of spare parts and Baxter is unable to replace the software with an alternate version. As a result, Baxter requested the accounts to remove the pumps from service and return them to Baxter. The accounts were requested to call Baxter''s Medication Delivery Services at 1-800-843-7867 to arrange for the return of the units and replacement with 6060 Multi-Therapy Infusion Pumps.

Distribution

Nationwide including Puerto Rico, and intenationally to Canada, China, Belgium, Hong Kong and Saudi Arabia