Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.
Recall
- Recall Number
- Z-1592-05
- Event Number
- 33032
- Firm
- Baxter Healthcare Corp. Rt.
- FEI Number
- 1416980
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 17, 2005
- Posted
- September 28, 2005
- Terminated
- June 19, 2007
- Address
- 120 & Wilson Rd, Round Lake, IL, 60073
Description
Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased by Baxter Healthcare Corp., Deerfield, IL in January 2000.
Potential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher
Baxter sent the 8/17/05 Urgent Product Recall letter to all Baxter 6060E Infusion Pump customers to inform them of a software anomaly present in software versions 3.00 and higher, which allows the pump to overinfuse in the Auto-Ramp mode. The end users were also informed that these epidural pumps, identified by the yellow housing, can no longer be supported due to a lack of availability of spare parts and Baxter is unable to replace the software with an alternate version. As a result, Baxter requested the accounts to remove the pumps from service and return them to Baxter. The accounts were requested to call Baxter''s Medication Delivery Services at 1-800-843-7867 to arrange for the return of the units and replacement with 6060 Multi-Therapy Infusion Pumps.
Nationwide including Puerto Rico, and intenationally to Canada, China, Belgium, Hong Kong and Saudi Arabia