FDA Recall Open, Classified

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Recall: Z-1591-2026 · Initiated February 16, 2026

Recall

Recall Number
Z-1591-2026
Event Number
98393
Firm
Abiomed, Inc.
FEI Number
1220648
Product Code
OZD
Status
Open, Classified
Root Cause
Software design
Initiated
February 16, 2026
Posted
March 27, 2026
Address
22 Cherry Hill Dr, Danvers, MA, 01923-2575

Description

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Reason

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

Action

Abiomed notified international consignees on about 02/16/2026 via emailed CRITICAL PRODUCT CORRECTION letter. Consignees were instructed to cease use of the Automated Impella Controller (AIC) with an Impella 5.5 S1 (product code 0550-0007) unless AIC software is updated to v10.1 or above, identify current software in the lower left corner of the initial AIC display screen, contact your local Abiomed representative to schedule your software update, review and complete the customer response form provided and post this notification in a visible area for awareness. Consignees were also instructed to forward this notification to anyone within the facility that should be aware and forward the notification to consignees if further distributed.

Distribution

International distribution in the country of Australia, Canada, France, Germany, Kuwait.

Quantity

62 units