Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. Impella RP US Pump Set; Product Code: 004334. 3. Purge Cassette, Packaged; Product Code: 0043-0002. Also included inside Product Code 0043-0003 above. Product Code 0043-0002 also includes innermost packaged product with Product Code 0043-0001. 4. Impella 2.5 Set; Product Code: 004413 (Customer facing code 005040). 5. Impella RP Pump Set, EU; Product Code: 0046-0011. 6. Impella CP Pump Set, EU; Product Code: 0048-0002. 7. Impella CP Pump Set BR; Product Code: 0048-0002-BR. 8. Impella CP Smart Assist Set, EU; Product Code: 0048-0014 (Also inclusive of codes 0048-0047 and 1000402 listed below). 9. Impella CP Smart Assist Set, Canada; Product Code: 0048-0044. 10. Impella CP Smart Assist Set APAC; Product Code: 0048-0047 (inclusive under code 0048-0014). 11. Impella 5.0 IMC Pump Set EU; Product Code: 005060. 12. Impella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002. 13. Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004. 14. 5.5 Accessories AU; Product Code: 1000362. 15. Impella CP Smart Assist Set; Product Code: 1000402 (inclusive under code 0048-0014).
Recall
- Recall Number
- Z-1589-2026
- Event Number
- 98395
- Firm
- Abiomed, Inc.
- FEI Number
- 1220648
- Product Code
- OZD
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- February 18, 2026
- Posted
- March 31, 2026
- Address
- 22 Cherry Hill Dr, Danvers, MA, 01923-2575
Description
Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. Impella RP US Pump Set; Product Code: 004334. 3. Purge Cassette, Packaged; Product Code: 0043-0002. Also included inside Product Code 0043-0003 above. Product Code 0043-0002 also includes innermost packaged product with Product Code 0043-0001. 4. Impella 2.5 Set; Product Code: 004413 (Customer facing code 005040). 5. Impella RP Pump Set, EU; Product Code: 0046-0011. 6. Impella CP Pump Set, EU; Product Code: 0048-0002. 7. Impella CP Pump Set BR; Product Code: 0048-0002-BR. 8. Impella CP Smart Assist Set, EU; Product Code: 0048-0014 (Also inclusive of codes 0048-0047 and 1000402 listed below). 9. Impella CP Smart Assist Set, Canada; Product Code: 0048-0044. 10. Impella CP Smart Assist Set APAC; Product Code: 0048-0047 (inclusive under code 0048-0014). 11. Impella 5.0 IMC Pump Set EU; Product Code: 005060. 12. Impella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002. 13. Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004. 14. 5.5 Accessories AU; Product Code: 1000362. 15. Impella CP Smart Assist Set; Product Code: 1000402 (inclusive under code 0048-0014).
Increased risk of purge leaks with Generation 1 purge cassettes.
Abiomed notified consignees on about 02/18/2026 via UPS next day airmailed letter titled "URGENT MEDICAL DEVICE RECALL (REMOVAL)." Consignees were instructed to identify and quarantine any affected Purge Cassettes in inventory, utilize the provided return shipment label to initiate return of affected cassettes, and complete and return the business reply form provided. Abiomed will provide a credit memo for returned units. Consignees were also encouraged to follow their standard process when ordering their Generation 2 Purge Cassettes, forward the notification to anyone within the facility needing to be aware as well as any customers if the product was further distributed, and to post the notification in a visible area for awareness.
Worldwide distribution.
33,107 units