Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placement, embolization and thrombolysis. Cardiac imaging applications including diagnostics, interventional minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper ED series is compatible with a hybrid operating room.
Recall
- Recall Number
- Z-1589-2018
- Event Number
- 79770
- Firm
- Philips Electronics North America Corporation
- FEI Number
- 1218950
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 14, 2018
- Terminated
- April 29, 2024
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placement, embolization and thrombolysis. Cardiac imaging applications including diagnostics, interventional minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper ED series is compatible with a hybrid operating room.
After continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.
Field Safety Notices were distributed on 3/14/18. The notices instructed customers: To prevent occurrence of the issue customers shall restart the system at least once a day. Customer shall ensure that all staff with access to the affected systems are informed of the contents of this Field Safety Notice. A copy of this Field Safety Notice shall be placed together with the documentation of the system until the system has been corrected by Philips. The problems will be resolved by a software update, which is expected to be available by the second half of 2018. You will be notified by your local Philips representative when the software update is available for installation. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377, 8am to 8pm (EST) Monday thru Friday.
Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The product was distributed to the following countries: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Croatia, Cuba, Czech Republic, Denmark, Estonia, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Republic of Korea, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, Norway, Pakistan, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, and United Kingdom.
682