Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
Recall
- Recall Number
- Z-1588-2013
- Event Number
- 65137
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 21, 2013
- Posted
- June 25, 2013
- Terminated
- November 27, 2013
- Address
- 595 Miner Rd, Cleveland, OH, 44143-2131
Description
Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse. Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol
Philips Medical Systems sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated Mar 21, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Nationwide Distribution-including the states of AR, AZ, DE, IN, KS, LA, MA, MN, MS, MT, NJ, NY, NY, OH, OR, PA, TN, TX, VT, WV, and WY.
50 units