FDA Recall Terminated

Analyst MD Version 1.6.1 and 1.6.2 Software used with the following instruments: API 3200MD" LC/MS/MS System, Instrument Part Number: 5024501; 3200MD QTRAP LC/MS/MS System, Instrument Part Number: 5024500; Triple Quad" 4500MD LC/MS/MS System, Instrument Part Number: 5031257; QTRAP 4500MD LC/MS/MS System, Instrument Part Number: 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.

Recall: Z-1586-2016 · Initiated February 24, 2016

Recall

Recall Number
Z-1586-2016
Event Number
73433
Firm
Ab Sciex
FEI Number
3008935545
Product Code
DOP
Status
Terminated
Root Cause
Software design
Initiated
February 24, 2016
Posted
April 29, 2016
Terminated
February 6, 2017
Address
500 Old Connecticut Path, Framingham, MA, 01701-4574

Description

Analyst MD Version 1.6.1 and 1.6.2 Software used with the following instruments: API 3200MD" LC/MS/MS System, Instrument Part Number: 5024501; 3200MD QTRAP LC/MS/MS System, Instrument Part Number: 5024500; Triple Quad" 4500MD LC/MS/MS System, Instrument Part Number: 5031257; QTRAP 4500MD LC/MS/MS System, Instrument Part Number: 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.

Reason

Wrong quantitative results may be displayed in a report from the device, which may potentially lead to an incorrect patient diagnosis.

Action

Sciex sent an Urgent Medical Device Correction letter dated February 29, 2016, to all affected US customers via Fedex on March 2, 2016. International customers were notified beginning on March 3, 2016. The letter informed customers of the issue and requests that they take temporary steps to eliminate the potential for erroneous events. In addition, a software update is in development and is expected to be available in approximately 4 weeks. Customers with questions were instructed to contact SCIEX at 1-289-982-2531.

Distribution

Worldwide Distribution - US (Nationwide) and Internationally to the following countries: AUSTRIA CANADA CHINA FRANCE GERMANY INDIA IRELAND ITALY KOREA MALAYSIA NETHERLANDS NORWAY RUSSIA SINGAPORE SPAIN SWITZERLAND UNITED KINGDOM

Quantity

279