ACUSON SC2000 ultrasound systems at software versions VA20A or VA20B. Model number 10433816. The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications.
Recall
- Recall Number
- Z-1584-2015
- Event Number
- 71122
- Firm
- Siemens Medical Solutions USA, Inc.
- FEI Number
- 2936884
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 11, 2012
- Posted
- May 6, 2015
- Terminated
- May 6, 2015
- Address
- 685 E Middlefield Rd, Mountain View, CA, 94043-4045
Description
ACUSON SC2000 ultrasound systems at software versions VA20A or VA20B. Model number 10433816. The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications.
In certain situations, loss of post exercise images can occur in Stress Echo.
On 07/11/2012 a field corrective action was initiated via update programs US016/12/P to update all affected SC2000 customers to software version VA20C, which contains the fix to the described issue. Each customer received a site visit to install the software update.
Worldwide Distribution -- US and Albania, Australia, Austria, Bangladesh, Bulgaria, Canada, Colombia, Denmark, Ecuador, Egypt, France, Germany, Hungary, Iceland, India, Israel, Italy, Japan, Kenya, Kuwait, Malaysia, Mauritius, Mexico, Netherlands, P.R. China, Philippines, Portugal, Rep. of Yemen, Republic Korea, Russian Fed., San Marino, Saudi Arabia, Singapore, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, and Venezuela.