FDA Recall
Terminated
Cardiomedics CardiAssist External Counterpulsation System, circulatory assist device, Series 4000.
Recall: Z-1582-2009
·
Initiated April 9, 2009
Recall
- Recall Number
- Z-1582-2009
- Event Number
- 52223
- FEI Number
- 2025845
- Product Code
- DRN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 9, 2009
- Posted
- July 7, 2009
- Terminated
- August 27, 2012
- Address
- Cardiomedics Incorporated 7 Whatney Suite B, Irvine, CA, 92618-2839
Description
Cardiomedics CardiAssist External Counterpulsation System, circulatory assist device, Series 4000.
Reason
Cuff may underinflate, or device may stop working during use. Cardioassist software version 4.3.1 utilized a signal from the plethysmograph that was found not to be sensitive enough due to the shape of the waveform in certain circumstances where there was excessive motion artifact.
Action
Service visit to customer in May 2009 to upgrade and validate CardioAssist 4000 software to version 4.3.5.
Distribution
Florida
Quantity
1