FDA Recall Terminated

TumorLoc software application, Software Versions: 23.1.1, 20.7.13, and 20.8.15 for use on the GEMINI TF 16 and GEMINI TF 64 Slice PET/CT Systems, Catalog Numbers: 882470 - GEMINI TF 16 Slice; 882471 - GEMINI TF 64 Slice; and 882473 - GEMINI TF 16 Slice Base Configuration. Model Numbers: 4535 679 83931 - GEMINI TF 16 Slice; 4535 679 94741 - GEMINI TF-64 Slice and 4535 674 41711 - GEMINI TF 16 Slice Base Configuration. Tumor Localization is an application used in radiation oncology departments for localizing target volumes for radiation therapy planning. It includes exclusive features for viewing Respiratory Correlated CT data sets and analyzing motion of target and surrounding anatomy.

Recall: Z-1578-2009 · Initiated February 19, 2009

Recall

Recall Number
Z-1578-2009
Event Number
52176
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
February 19, 2009
Posted
July 6, 2009
Terminated
September 21, 2012
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

TumorLoc software application, Software Versions: 23.1.1, 20.7.13, and 20.8.15 for use on the GEMINI TF 16 and GEMINI TF 64 Slice PET/CT Systems, Catalog Numbers: 882470 - GEMINI TF 16 Slice; 882471 - GEMINI TF 64 Slice; and 882473 - GEMINI TF 16 Slice Base Configuration. Model Numbers: 4535 679 83931 - GEMINI TF 16 Slice; 4535 679 94741 - GEMINI TF-64 Slice and 4535 674 41711 - GEMINI TF 16 Slice Base Configuration. Tumor Localization is an application used in radiation oncology departments for localizing target volumes for radiation therapy planning. It includes exclusive features for viewing Respiratory Correlated CT data sets and analyzing motion of target and surrounding anatomy.

Reason

A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software defect results in a the potential for data to be lost during reconstruction of images in which slices can either be discarded, or interleaved based upon their series ID.

Action

An "URGENT - Field Safety Notice" dated February 12, 2009 was sent via FedEx to domestic customers. The notice described the affected products, issue, involved hazard should problem occur and corrective actions for the customer/user. The letter also notified users/customers that a Philips Healthcare Field Service Engineer will contact and visit their site within 6 months to install a TumorLoc upgrade. Notification will also be sent to the international consignees by Philips personnel. For further information or support concerning this issue, please contact your local Philips representative: Customer Care Center (1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: Oncology) or to your local Philips Healthcare office.

Distribution

Worldwide Distribution -- US, Columbia, Germany, France, Netherlands, Canada, Belgium and Slovenia.

Quantity

13 units