Regard, Item Number: 880342003, Sterile, GS0776C - Laparotomy - OSU
Recall
- Recall Number
- Z-1577-2017
- Event Number
- 76255
- Firm
- Resource Optimization & Innovation Llc
- FEI Number
- 3004513970
- Product Code
- LRO
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 9, 2017
- Terminated
- September 28, 2018
- Address
- 2909 N Neergard Ave, Springfield, MO, 65803-6317
Description
Regard, Item Number: 880342003, Sterile, GS0776C - Laparotomy - OSU
Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.
ROi sent an Urgent Recall Notice dated January 9, 2017, to all affected consignees. The recalling firm instructed consignees to quarantine affected product and complete the response form. The recalling firm stated that they will issue over lableing for the affected packs. Sterile replacement light covers will be sent directly to the end user if available. Consignees were instructed to inform their subaccounts of the recall. For questions regarding this recall call 417-820-2793.
US Distribution to LA and TX.
78 kits