FDA Recall Terminated

XoranConnect, software version 4.1.0.6, Xoran Technologies, Ann Arbor, MI. A web-based software which provides online multi-planar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. Server designed to store and serve stored imaging content on a RAID5 storage device.

Recall: Z-1577-2009 · Initiated February 3, 2009

Recall

Recall Number
Z-1577-2009
Event Number
52171
Firm
Xoran Technologies, Inc.
FEI Number
3004198450
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
February 3, 2009
Posted
July 9, 2009
Terminated
November 19, 2009
Address
309 N 1st St, Ann Arbor, MI, 48103-3301

Description

XoranConnect, software version 4.1.0.6, Xoran Technologies, Ann Arbor, MI. A web-based software which provides online multi-planar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. Server designed to store and serve stored imaging content on a RAID5 storage device.

Reason

An incorrect patient name label may appear on the screen displaying the selected patient's scan.

Action

The firm released software version 4.1.0.7 by uploading it through a remote software upgrade on February 3, 2009. User accounts were notified by letter dated February 9, 2009. Direct questions to Xoran Technologies at 1-800-709-6726.

Distribution

Nationwide Distribution (Alabama, Arizona, Colorado, Georgia, Illinois, Louisiana, New York, North Carolina, Oklahoma, Rhode Island, Tennessee, Texas and Utah).

Quantity

14