Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL, and a lysing reagent 125 mL.
Recall
- Recall Number
- Z-1576-2016
- Event Number
- 73650
- Firm
- Jas Diagnostics/Drew Scientific
- FEI Number
- 3003071743
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- March 10, 2016
- Terminated
- August 21, 2019
- Address
- 14100 NW 57th Ct, Miami Lakes, FL, 33014-3107
Description
Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL, and a lysing reagent 125 mL.
The reagent pack may result in elevated platelet background during Drew3 Hematology Analyzer routine check.
JAS Diagnostics/ Drew Scientifics sent an Important Product Advisory letter dated March 10, 2016, to all affected customers. The domestic customers were notified via letter and international customers were notified via email. Customers were instructed to contact Drew Scientific Inc. Customer Service or Technical Service at 1-800-433-0945 for product replacement. The firm suggested a temporary workaround of performing five background runs three times per day in order to ensure background counts are running within specification.
Worldwide Distribution - US Distribution including the states of : AL, AR, CA, FL, GA, IA, IN, KS, KY, LA, MA, MN, MO, MS, NC, NE, NY, SC, TN, and TX. and to the countries of : Bolivia, Cairo, Columbia, Ecuador, Italy, Moscow, Netherlands, Pakistan, Peru, Slovenia, Spain, and Venezuela.