FDA Recall Terminated

ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System.

Recall: Z-1576-05 · Initiated March 9, 2004

Recall

Recall Number
Z-1576-05
Event Number
32608
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
MZO
Status
Terminated
Root Cause
Other
Initiated
March 9, 2004
Posted
October 1, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System.

Reason

An error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek ELx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay.

Action

Dear Customer letters, dated 03/08/2004 were sent to customers informing them of the specification error and instructing users to validate their assays manually.

Distribution

AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI & WV

Quantity

128