FDA Recall Terminated

Regard, Item Number: 880331002, Sterile, OR0762B - Total Knee - OSU

Recall: Z-1573-2017 · Initiated January 9, 2017

Recall

Recall Number
Z-1573-2017
Event Number
76255
Firm
Resource Optimization & Innovation Llc
FEI Number
3004513970
Product Code
LRO
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 9, 2017
Terminated
September 28, 2018
Address
2909 N Neergard Ave, Springfield, MO, 65803-6317

Description

Regard, Item Number: 880331002, Sterile, OR0762B - Total Knee - OSU

Reason

Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.

Action

ROi sent an Urgent Recall Notice dated January 9, 2017, to all affected consignees. The recalling firm instructed consignees to quarantine affected product and complete the response form. The recalling firm stated that they will issue over lableing for the affected packs. Sterile replacement light covers will be sent directly to the end user if available. Consignees were instructed to inform their subaccounts of the recall. For questions regarding this recall call 417-820-2793.

Distribution

US Distribution to LA and TX.

Quantity

8 kits