FDA Recall Terminated

NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Recall: Z-1573-2016 · Initiated March 24, 2016

Recall

Recall Number
Z-1573-2016
Event Number
73741
Firm
Philips and Neusoft Medical Systems Co., Ltd. No. 2 Xiu xiu Street Hunnan; Sujiatun Shenyang China
FEI Number
3003575350
Product Code
JAK
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 24, 2016
Posted
April 27, 2016
Terminated
November 23, 2016

Description

NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Reason

The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be

Action

Affected consignees will be notified of the recall via letter on March 24, 2016. Mandatory Field Change Order and FCO kits will be released to all affected systems in June, 2016.

Distribution

NC, OH, NE, SC, TX, LA, PR, MO, FL, CT

Quantity

37 units