Medtronic SynchroMed II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.
Recall
- Recall Number
- Z-1570-2014
- Event Number
- 67720
- Firm
- Medtronic Neuromodulation
- FEI Number
- 2182207
- Product Code
- LKK
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- February 26, 2014
- Posted
- May 8, 2014
- Terminated
- September 28, 2018
- Address
- 7000 Central Ave NE, Minneapolis, MN, 55432-3568
Description
Medtronic SynchroMed II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.
This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re
Medtronic sent a "Urgent Medical Device Correction" letter dated March 2014. The letter was sent to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter provided the Explanation of the Issue, Scope and Severity, Recommendations, and Important Guidelines. Customer visits were started by Medtronic field Representatives on February 26th, 2014. Medtronic is communicating this information to the appropriate regulatory agencies globally, including the U.S. Food and Drug Administration. We are committed to continuing to improve our product performance and services to enable you to manage your patients in a safe and effective manner. If you have questions, please contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7am- 6pm CST.
Worldwide Distribution - All states in USA. OUS: List not provided at this time.
195,198 pumps (146,435 US, 48,763 OUS)