FDA Recall Terminated

Sterrad 50 Sterilizer, Product Code 10050. Low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Recall: Z-1570-2009 · Initiated January 12, 2009

Recall

Recall Number
Z-1570-2009
Event Number
52124
Firm
Advanced Sterilization Products
FEI Number
3003662624
Product Code
MLR
Status
Terminated
Root Cause
Other
Initiated
January 12, 2009
Posted
June 30, 2009
Terminated
December 28, 2011
Address
33 Technology Drive, Irvine, CA, 92618

Description

Sterrad 50 Sterilizer, Product Code 10050. Low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Reason

The capacitor in the vacuum pump component within the Sterrad 50 System could overheat and potentially create a hazardous condition inside the system, resulting in smoke and/or fire.

Action

Advanced Sterilization Products (ASP) sent an "Urgent: Product Correction" letter dated January 12, 2009 to each consignee informing them of the issue and instructed them to report smoke or a burning odors in or around their STERRAD 50 System. ASP strongly recommended for consignees to follow their facility's safety protocols and if possible, immediately disconnect the STERRAD 50 System from its electrical source. Customers were instructed to contact ASP to schedule an onsite visit to investigate the issue. If the issue is caused by a malfunctioning capacitor in the vacuum pump, the firm will replace the malfunctioning component right away. To correct this situation, an ASP Field Service Engineer (FSE) will visit all customers facilities within the next sixty days to replace the capacitor in the vacuum pump on the affected STERRAD 50 System.

Distribution

Worldwide Distribution -- US, Germany, Italy, France, United Kingdom, Switzerland, Benelex, Uruguay, Venezuela, Argentina, Chile, Columbia, Belgium, Puerto Rico, Mexico, Philipines, Australia, Japan, Korea, Malaysia, Taiwan, Middle East, Peru, Canada, Czech Republic, Portugal, Spain, Sweden, Hungary, Poland, Brazil, Russia, Finland, Norway, Panama and Switzerland.

Quantity

1,151 units.