Skytron Ergon 3 Series Skyboom Monitor Bracket-Flatscreen Bracket, single monitor mount, 27" max. screen size Part number 3FCM1

Recall: Z-1569-2020 · Initiated March 5, 2020

Recall

Recall Number
Z-1569-2020
Event Number
85252
Firm
Skytron, Div. The KMW Group, Inc
FEI Number
1825014
Product Code
FQO
Status
Open, Classified
Root Cause
Device Design
Initiated
March 5, 2020
Address
5085 Corporate Exchange Blvd SE, Grand Rapids, MI, 49512-5515

Description

Skytron Ergon 3 Series Skyboom Monitor Bracket-Flatscreen Bracket, single monitor mount, 27" max. screen size Part number 3FCM1

Reason

Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.

Action

Skytron customer notification letters mailed on 3/5/2020. Distributors received notification via email on 3/5/2020. Letter states reason for recall, health risk and action to take: 1.Identify and inspect your 3FCM1 monitor bracket(s). 1508T719-00-0-0120 2.Remove from service any bracket with noticeable physical distortion or cracking around the welded area. 3.Skytron authorized representatives have been directed to contact and schedule a service visit within a 30-day window of time. The contact information for your authorized representative is: Johnsons Medical, Inc. Donna Johnson, (801) 768-4411 Please contact Erin Woolf, at 616-656-2998 or [email protected] with any questions. For after-hours support, please call 1-800-SKYTRON.

Distribution

AZ, CO, GA, IL, MI, MN, NJ, PA, VA, VT

Quantity

43 units