FDA Recall Terminated

Shimadzu Digital Radiography X-Ray System, Catalog No. DAR-8000f. This device is intended to be used for radiography in the hospital with X-ray devices.

Recall: Z-1569-2014 · Initiated April 3, 2014

Recall

Recall Number
Z-1569-2014
Event Number
67927
Firm
Shimadzu Medical Systems
FEI Number
1000125362
Product Code
IZI
Status
Terminated
Root Cause
Software design
Initiated
April 3, 2014
Posted
May 8, 2014
Terminated
June 25, 2015
Address
20101 S Vermont Ave, Torrance, CA, 90502-1328

Description

Shimadzu Digital Radiography X-Ray System, Catalog No. DAR-8000f. This device is intended to be used for radiography in the hospital with X-ray devices.

Reason

Shimadzu Corporation is recalling the Shimadzu Digital X-Ray System because they have received a report of a lost image under certain circumstances.

Action

A Service Bulletin and a customer notification letter were sent on 4/3/2014 for all the customers who purchased the Shimadzu Digital X-Ray System (Catalog No. DAR-8000f). The notification informs the customers of the problems identified and the actions to be taken. Customers with questions or concerns about the device correction, are instructed to contact their local Shimadzu representative or contact Shimadzu National Technical Support at (800) 228-1429.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Japan, China, Germany, Australia, Thailand, Sri Lanka, South Africa, UAE, Saudi Arabia, Korea, and Taiwan.

Quantity

685 units total (14 units in US)