FDA Recall Open, Classified

Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973

Recall: Z-1566-2023 · Initiated March 22, 2023

Recall

Recall Number
Z-1566-2023
Event Number
91874
Firm
ICU Medical Inc
FEI Number
3013319212
Product Code
FRN
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 22, 2023
Posted
May 16, 2023
Address
600 N Field Dr, Lake Forest, IL, 60045-4835

Description

Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973

Reason

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

Action

ICU medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 03/22/2023 to its direct and indirect consignees by mail. The notice explained the issue and the hazard and requested the following: Clinical Users: Whenever possible, keep the pump plugged into AC power. Before disconnecting the pump from AC power, please ensure that the battery is fully charged. Closely monitor the Battery Status Indicator while the pump is disconnected from AC power. Additionally, have a backup pump available when infusing critical medications. Biomedical Engineering: You may replace affected batteries with a new CSB battery until corrected batteries are available. Please do not use a replacement battery with corroded battery terminals. 1. Identify all affected batteries in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification and proposed mitigations. 2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected].

Distribution

worldwide

Quantity

23815 units