PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Model Numbers: 620-040-610, 620-040-611, 620-040-613, 620-040-614, 620-040-600, 620-040-601, 620-040-602, 620-040-604, 105-210-684 The Pneumo Sure XL High Flow Insufflator is a device that serves to create a cavity by insufflating CO2 during diagnostics and/or therapeutic laparoscopy. The insufflator consists of a console, that is intended to be connected to a CO2 gas source, and has a tubeset inserted to transport CO2 gas from the insufflator to the connecting trocar.
Recall
- Recall Number
- Z-1566-2016
- Event Number
- 73674
- Firm
- Stryker Endoscopy
- FEI Number
- 2936485
- Product Code
- HIF
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 28, 2016
- Terminated
- December 14, 2016
- Address
- 5900 Optical Ct, San Jose, CA, 95138-1400
Description
PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Model Numbers: 620-040-610, 620-040-611, 620-040-613, 620-040-614, 620-040-600, 620-040-601, 620-040-602, 620-040-604, 105-210-684 The Pneumo Sure XL High Flow Insufflator is a device that serves to create a cavity by insufflating CO2 during diagnostics and/or therapeutic laparoscopy. The insufflator consists of a console, that is intended to be connected to a CO2 gas source, and has a tubeset inserted to transport CO2 gas from the insufflator to the connecting trocar.
When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.
Urgent Medical Device Recall letters were sent to domestic customers on March 28, 2016 by certified mail. Stryker International distribution sites were notified and are responsible for providing information to affected customers.
Worldwide distribution. US nationwide including Puerto Rico, China, Hong Kong, India, Japan, Korea, Romania, Vietnam, Singapore, Taiwan, Netherlands, Peru, Australia, Canada, Mexico, Argentina, Chile, Poland, South Africa, and Guatemala.