Automated peritoneal dialysis (APD) cycler
Recall
- Recall Number
- Z-1565-2016
- Event Number
- 73435
- Firm
- Baxter Healthcare Corp.
- FEI Number
- 1417572
- Product Code
- FKX
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- February 16, 2016
- Terminated
- December 23, 2016
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
Automated peritoneal dialysis (APD) cycler
Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations. The tot
An Urgent Device Correction communication was sent to the one affected customer via U.S.P.S., first class mail on 02/11/2016. A Clinician Guide Addendum was created to instruct clinicians of the extra 200 mL of PD solution required to prime the patient line and for air purge operations. This Addendum was provided as an attachment to the Urgent Device Correction communication letter.
US Consignee: NY ***Foreign Consignee: Ecuador
10 units