3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.
Recall
- Recall Number
- Z-1563-2016
- Event Number
- 73117
- Firm
- 3M Company Health Care Business 3m Center
- FEI Number
- 2110898
- Product Code
- FLF
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 22, 2016
- Terminated
- August 25, 2016
- Address
- 2510 Conway Ave, B # 275-5-W-6, Saint Paul, MN, 55144-0001
Description
3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.
3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.
3M sent an Urgent Medical Device Field Correction letter to customers beginning 01/22/2016. The letter identified affected product, and stated that a 3M Service representative will contact the consignee to inspect the sterilizer. Questions can be directed to 3M Health Care Service Line at 1-800-228-3957.
AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include: Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.
74 valves