FDA Recall Terminated

GE Healthcare Precision RXi 23A/ 32A Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, 65 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of exams for various clinical applications with ether 23 cm image intensifier with 512x512 Image Chain or 32 cm Image intensifier with either 512x512 Image Chain or 1Kx 1K Image Chain.

Recall: Z-1563-2008 · Initiated December 18, 2007

Recall

Recall Number
Z-1563-2008
Event Number
46807
Firm
GE Healthcare
FEI Number
2126677
Product Code
JAA
Status
Terminated
Root Cause
Device Design
Initiated
December 18, 2007
Posted
August 24, 2008
Terminated
August 9, 2012
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Precision RXi 23A/ 32A Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, 65 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of exams for various clinical applications with ether 23 cm image intensifier with 512x512 Image Chain or 32 cm Image intensifier with either 512x512 Image Chain or 1Kx 1K Image Chain.

Reason

GE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the remote console joystick where a fatal accident occurred on a non-GE Remote R&F System. .

Action

Two letters "GE Product Safety Notification" dated 12/18/07 and 3/7/08, addressed to Hospital Administrators, Manager of Radiology and Radiologists, were sent to each consignee. Each letter addressed a separate safety issue. Letters described the affected product, safety instructions, corrective action and contact information. The joystick can become stuck causing unintentional motion of the tube. Field Modification Instructions (FMI 10802/10803) have been issued to the field to install additional warning labels on the patient table, install a protector for the console joystick and inspect and correct the clearance between the console joystick and its mounting location. If your facility has a Precision RXi or Precision e system manufactured before March 2007, a GE Healthcare Field Engineer will schedule a field visit to install the added joystick protection. This upgrade will be performed at no charge. If you have any questions/concerns regarding this issue, please contact your local GE Healthcare sales or service representative.

Distribution

(USA) AK CO IL LA MA MI MS NC NJ NY PA SC TN UT VA WA WI (OUSA) BR CA MX AU C3 KR PH TW AE AT BE BF BG CM CY CZ DE ES FI FR GB GE GR GY HR HU IT LB LV MT NA NG PL PT RO RU SA SD SE SI SK SY TN TR UA YU ZA

Quantity

66