GE Signa Ovation 0.35 T MR System. The Signa Ovation System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times.
Recall
- Recall Number
- Z-1556-2009
- Event Number
- 52015
- Firm
- GE Medical Systems, LLC
- FEI Number
- 2126677
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 16, 2009
- Posted
- June 30, 2009
- Terminated
- April 23, 2013
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Signa Ovation 0.35 T MR System. The Signa Ovation System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times.
GE Healthcare has identified a potential safety issue with the Signa OpenSpeed and Signa Ovation MR Scanner tables that may impact patient safety. The table may begin lowering on its own while a patient is on the table. If this occurs while the table is in the magnet bore, the table may tilt and could result in patient injury. The patient's head or feet depending on patient orientation, may hit
GE Healthcare issued an "Urgent Medical Device Correction" letter dated April 16, 2009 addressed to Hospital Administrators, Mangers of Radiology/Cardiology and Radiologists/Cardiologists. The letter describes the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. A GE Healthcare Field Engineer will schedule a field visit to correct the device. Further questions or concerns may be addressed to GE Healthcare at 1-262-521-6681.
Worldwide Distribution -- US including PR and states of AL, AZ, AR, CA, CO,CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MN, MS, MO,MT, NV, NH, NE, NJ, NY, NC, OH, OK,OR, PA, SC, SD, TN, TX, VA, WA, WV, WI and countries of BRAZIL CHINA, DENMARK, EGYPT, ENGLAND, GERMANY, GUATEMALA, HUNGARY, ITALY, IVORY COAST, JAPAN, KOREA, KUWAIT, OMAN, MEXICO, OMAN, PHILIPPINES, RUSSIAN FEDERATION, SPAIN, SWITZERLAND, TURKEY, UNITED ARAB EMIRATES and VENEZUELA.