FDA Recall Terminated

Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Recall: Z-1555-2011 · Initiated August 23, 2007

Recall

Recall Number
Z-1555-2011
Event Number
57688
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IYE
Status
Terminated
Root Cause
Other
Initiated
August 23, 2007
Posted
March 4, 2011
Terminated
April 7, 2011
Address
4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011

Description

Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Reason

Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent within the iCom tolerance table.

Action

The firm, Elekta, sent an "Important Notice A302" dated August 23, 2007 titled "Update to iCom Tables for Desktop Pro when used with Third Party R&V Systems",to all affected customers. On September 28, 2007, Elekta issued a second Important Notice A304 which reiterated the same information as was in Important Notice A302. The notices described the product, problem and actions to be taken. This notice provides a workaround on how to avoid the issue. In order to avoid this issues documented, Elekta recommended that the customers set the tolerance for collimators in the iCom tolerance table to be 3.0 degrees-no less than 3.0 and to use the Linac Record, if configured, as a means of checking the final collimator position. Elekta noted that the issue will be resolved in a future release of Desktop Pro. If you any questions related to this issue, please contact your local Elekta, third party R&V system representative or Elekta Limited at +44 (0)1293 654200.

Distribution

Worldwide distribution: USA including Puerto Rico and countries including: Canada and Mexico.

Quantity

229 units