FDA Recall Terminated

Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the cardiac troponin I method on the Stratus CS analyzer.

Recall: Z-1555-2009 · Initiated April 28, 2008

Recall

Recall Number
Z-1555-2009
Event Number
52007
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
MMI
Status
Terminated
Root Cause
Labeling False and Misleading
Initiated
April 28, 2008
Terminated
July 14, 2009
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the cardiac troponin I method on the Stratus CS analyzer.

Reason

The firm received and confirmed that some lots of Cardiac Troponin I (cTnl) CalPaks (CCTNI-CR) may exhibit calibration issues that cause lower than typical signal during calibration with the current population of cTnl TestPaks (CCTNI). This issue may result in failed calibrations due to "Slope Error" messages or falsely elevated patient sample and quality control (QC) results on calibrations that

Action

Siemens Healthcare Diagnostics, Inc. issued an Urgent Field Safety Notice dated April 29, 2008 to all affected customers instructing them to discontinue use of 3 CalPak lots and reduce the shelf-life for the remaining lots. For further information, contact Siemens Healthcare Diagnostics, Inc. at 1-302-631-7672.

Distribution

Worldwide Distribution - including United States, United Arab Emirates, Austria, Bermuda, Belize, Canada, Germany, Egypt, India, Japan, Saudi Arabia and Bengal.

Quantity

4415 cartons