FDA Recall
Terminated
Waste bottles Part Numbers/Article Numbers: 905-590/D512 and 905-802/D513. The waste bottles are used to collect waste from the ABL7XX and the ABL8XX blood gas analyzer after rinsing.
Recall: Z-1554-2009
·
Initiated April 16, 2009
Recall
- Recall Number
- Z-1554-2009
- Event Number
- 51986
- Firm
- Radiometer America Inc
- FEI Number
- 1523456
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- April 16, 2009
- Posted
- July 15, 2009
- Terminated
- May 9, 2012
- Address
- 810 Sharon Dr, Westlake, OH, 44145
Description
Waste bottles Part Numbers/Article Numbers: 905-590/D512 and 905-802/D513. The waste bottles are used to collect waste from the ABL7XX and the ABL8XX blood gas analyzer after rinsing.
Reason
When removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.
Action
Radiometer America Inc. issued a recall notification letter dated May 5, 2009 to all customers containing information on replacement of all potentially defective waste containers and instruction on how to handle any mounted waste containers. For further questions, contact Radiometer America Inc. at 1-800-736-0600.
Distribution
Worldwide Distribution to US and Canada.
Quantity
905-590 (D512): 35,552 bottles; 905-802 (D513): 45,204 bottles